18 A handbook on GC Q Exactive orbitrap use
18.1 Acquisition parameters
Check: Analytical performance of the various acquisition modes in Orbitrap MS and MS/MS. https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/full/10.1002/jms.4195
From: https://analyticalsciencejournals.onlinelibrary.wiley.com/doi/full/10.1002/rcm.8609?campaign=wolacceptedarticle Orbitrap technology is based on ion oscillation frequency measurements around a central electrode, detected on a receiver plate.26 The mass spectrum is obtained after Fourier transformation of the oscillation signal with a two-point calibration. The longer the signal acquisition time, the better is the resolution.27 During this time, ions generated in the EI source are stored in an IT, the C-TRAP. The storage of ions in the C-TRAP plays a vital role in prematurely analysing discontinuous ions produced continuously by the chromatographic separation and the ionisation source. The ions introduced into the C-TRAP lose their energy in contact with N2 at approximately 1 mTorr.26 During an acquisition in full scan mode, it is possible to set the detector with different parameters: acquisition mass range, resolution, AGC target and tune values. The tune files gather the different settings of the mass spectrometer aiming to produce the ions of the EI source, but also the different voltages applied to the lenses in the ion path from the source to the C-TRAP. It is also possible to change the value of the C-TRAP energy offset (−5 to +5 V). For this study, we used standard conditions of EI (70 eV) in order to have the ideal conditions for reproducibility of the spectra. By default, the Orbitrap system acquires data in the range m/z 50–500 at 60 000 resolution mode and 70 eV for EI. The C-TRAP tune value was set to 0 V and AGT target to 1 × 106 ions with an automatic filling limit.
- Recommended to have about 10-12 scans per peak.
18.2 Chemical ionization
If the instrument is in CI mode, choose the right Calibration gas and Ionization mode (“CI”), as well as the CI gas flow (typically 1-2), gas port (Port B as of 20210429, actual CI gas flow should indicate the correct port), and gas type (likely Methane, but check the cylinder). Open the CI gas flow at 2ml/min before opening anything else (before the calibration gas, filament, MS), otherwise the flow can be unstable. Afterwards the CI can be adjusted to lower flow
18.3 Terminologies
Automatic Gain Control (AGC): Sets the ion injection time to maintain the optimum quantity of ions for each scan. With AGC on, the scan function consists of a prescan and an analytical scan. The split lens is used to start and stop the injection of ions into the mass analyzer. It provides a high deflection voltage most of the time so that ions are deflected into a baffle except when they are to be allowed into the C-Trap. The fast switching of the ion beam ensures the precise determination of the ion injection time that is required for AGC. This is since if too many ion are in the ion cloud going in to the orbitrap, then it will oversaturate the detector.
Chemical Ionization (CI):
Nitrogen gas: Nitrogen gas is used for the purging of the ion source when removing the ion volume. It is also used in the C-Trap, when the ions are accumulated and their energy is dampened using a bath gas (nitrogen).
18.4 Cleaning source
https://www.sisweb.com/referenc/tips/msclean.htm
18.5 Performance mixture
QC standard mix (100 fg OFN, 10 pg HCB, 10 pg DDT and 10 pg Endrin)
In NCI, OFN (100 fg) should give you around 800 000 – 1 000 000 area counts with an optimal system. This can vary based on the tune parameters and column condition.
Alkane mix C7-C40
Odd chain alkane mix C11-C25 to help find peaks in the C7-C40 mixture
Check if 4,4’-Dibromooctafluorobiphenyl can be used in PCI/NCI
Alkane only visible in PCI so if using CI then run alkane mix in PCI even for NCI runs.